BASIC ANESTHESIA MACHINE SERVICE

We at Summit Medical Equipment feel that routine service is an important part of ensuring patient and O.R. staff safety. The following is a list of tasks we will perform during a basic system service.

1. LEAK TEST HI PRESSURE
     A. O2 / N2O / Med Air
          1. Includes quick disconnects, fittings, ferrules
          2. Includes hoses, regulators, tanks, valves
               a. H-Tank
               b. E-Tank

2. LEAK TEST LOW PRESSURE
     A. Flowmeter to Common Outlet
          1. 100mm Hg Pressure/Less than 20mm Hg Loss in 30 seconds

3. LEAK TEST/OUTPUT TEST VAPORIZER
     (Precision/Out of Breathing Circle)
     A. Leak Test
          1. Closed Position
               a. 100mm Hg. Pressure/Less than 20mm Hg. Loss in
               30 seconds
          2. Closed Position
               b. 200mm Hg. Pressure/Less than 4mm Hg. Loss in
               30 seconds
     B. Output Test
          1. Riken or Infrared Spectrophotometer Analysis of Vaporizer
              Efficacy
          2. Tolerance 15% of Dial
          3. Written Report as to Status

4. LEAK TEST REBREATHING HEAD
     A. Inspect Seals, "O" Rings, Gaskets, Connections
     B. Inspect Pop-off Valve
          1. Leak Test: Internal
          2. Leak Test: External

5. CLEAN, LUBRICATE, TIGHTEN, ALIGN, LOCKTITE, TEFLON, ADJUST AS NECESSARY

6. REPLACE 3/16 INCH, 1/4 INCH, AND/OR 5/16 INCH I.D. CONDUCTIVE TUBING

7. INSTALL NEW (DISPOSABLE) ADULT & PEDI REBREATHING SETS

8. LEAK TEST AND INSPECT ACCESSORIES

     A. RB Bags & Sets
     B. NRB Systems
     C. Induction Chamber
     D. Face Masks
     E. Evac Tubing

9. INSPECT AND MAKE RECOMMENDATIONS ON
     A. O2 Supply System
     B. Waste Gas Evac System
     C. Vaporizer Servicing
     D. OSHA Issues

10. OTHER RECOMMENDED SERVICE ITEMS (at additional costs)
     A. Gaskets, Seals, Fittings, Quick Disconnects, Regulators,
          Adapters, Brackets as Needed
     B. Evac Kits, Tubing, F/Air Canisters (Activated Charcoal), F/Air
          Holders
     C. Professional Cleaning and Calibration of Precision Vaporizers
          (See Attached Vaporizer Service Protocol)
          1. Replace Wicks, Seals, "O" Rings
          2. Clean, Calibrate to Label
          3. Freight Handling Insurance To and From Service Center
          4. Loaner Vaporizer
          5. No Trip Charge to Return Vaporizer
          6. Vaporizer Exchange
 

 


basic service
Why service vaporizers?